OtherMOU
3.5F BETA-RAIL (TM) DELIVERY CATHETER
Received Apr 10, 2003 · Event occurred Mar 12, 2003
Report 1062385-2003-00007 · MDR key 453755
Device
Generic name
Catheter
Manufacturer
Novoste CorpModel number
P01814Catalog number
ABR-0340
Lot number
621791
Product problems
- Kinked
- Inaccurate Delivery
Patient
*
- Unknown (for use when the patient's condition is not known)
Narrative
Description of Event or Problem
MISADMINISTRATION. COMPLAINANT REPORTED THAT UPON DELIVERING 18.4GY SOURCE TRAIN TO THE LEFT CIRCUMFLEX (3.0 DIAMETER), THE 3.5F BETA-RAIL (TM) DELIVERY CATHETER KINKED RESULTING IN THE SOURCES NOT REACHING THE DESIGNATED TREATMENT SITE. THE ELAPSED TIME THE SOURCES WERE INSIDE THE VESSEL AT THE SITE WAS 12 SECONDS, RESULTING IN 6% OF 18.4 GY (EQUAL TO 1.1 GY DELIVERED TO THE PROXIMAL TREATMENT SITE).