inforMED
MalfunctionIOY

SUPPORT ARM 176

Received Feb 11, 2015 · Event occurred Jan 23, 2015

Report 8010042-2015-00048 · MDR key 4524235

Device

Catalog number

6405976

Product problems

  • Mechanical Problem

Patient

Not reported

  • No Patient Involvement

Narrative

Additional Manufacturer Narrative

THE PART HAS BEEN REQUESTED BACK FOR INVESTIGATION. A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED.

Additional Manufacturer Narrative

THE BROKEN SUPPORT ARM WAS RECEIVED BACK FOR INVESTIGATION, WHEREIN VISUAL INSPECTION OF THE SUPPORT ARM CONFIRMED THE REPORTED CRACKS. THE SUPPORT ARM WAS MANUFACTURED IN AUGUST OF 2002. THE POSITION OF THE BREAK INDICATES THAT THE SUPPORT ARM HAD VEEN OVERLOADED, HOWEVER, THE TRUE CAUSE FOR THE BROKEN SUPPORT ARM/ HAS/CAN NOT BE DETERMINED FROM THIS INVESTIGATION. THE SUPPORT ARM IS DESIGNED AND TESTED ACCORDING TO THE STANDARD EN 60601-01.

Description of Event or Problem

(B)(4).

Description of Event or Problem

IT WAS REPORTED THAT THE SUPPORT ARM HAD CRACKS. THERE WAS NO PATIENT INVOLVEMENT. (B)(4).