MalfunctionEGJ
GIBSON COOKE
Received Mar 27, 2003
Report 450137 · MDR key 450137
Device
Generic name
Sweat Test Apparatus
Manufacturer
Rodeo ProductsModel number
1PS.25ACatalog number
*
Lot number
*
Product problems
- Shock, electrical
- Grounding Malfunction
- Defective Component
Patient
5 YR
- Sweating
- Electric Shock
Narrative
Description of Event or Problem
DURING SWEAT CHLORIDE COLLECTION, SWEAT COLLECTION DEVICE MALFUNCTIONED ADMINISTERING AN ELECTRICAL SHOCK TO PT. INVESTIGATION REVEALED A DEFECTIVE TRANSISTOR CAUSED THE SHOCK.