inforMED
MalfunctionFSO

OPMI PENTERO

Received Feb 5, 2015 · Event occurred Jan 8, 2015

Report 9615010-2015-00001 · MDR key 4494613

Device

Generic name

Microscope, Surgical General & Plastic

Model number

NA

Catalog number

000000-0480-801

Lot number

NA

Product problems

  • Failure to Conduct
  • Electrical /Electronic Property Problem

Patient

NA

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

THE BIOMEDICAL ENGINEER REPORTED THAT DURING A CRANIAL SURGERY THE LIGHT ON THE PENTERO MICROSCOPE WENT OUT AND AN ERROR MESSAGE APPEARED. AFTER TRYING TO SWITCH TO THE MICROSCOPE BACKUP LAMP, THE ERROR MESSAGE "DEFECTIVE LAMP" APPEARED AND THERE WAS STILL NO LIGHT. A CALL TO THE MANUFACTURER'S TECHNICAL SUPPORT CONFIRMED THAT THE LIGHT SOURCE WAS NOT FUNCTIONING. THE DOCTOR DECIDED TO CLOSE INCISION WITHOUT COMPLETING THE SURGICAL PROCEDURE.

Remedial action

  • Repair