MalfunctionKDP
VAVD
Received Jan 29, 2015 · Event occurred Jul 18, 2013
Report 8010762-2014-01041 · MDR key 4461740
Device
Generic name
Regulator, Vacuum
Manufacturer
Maquet Cardiopulmoary AgModel number
70103.5905Catalog number
70103.5905
Product problems
- Unstable
- Device Operates Differently Than Expected
Patient
Not reported
- No Patient Involvement
Narrative
Description of Event or Problem
ON (B)(6) 2013 AT THE SSU AT MAQUET (B)(4) IN (B)(6) REPORTED THAT THE VACUUM WAS UNSTABLE ON THE VACUUM CONTROLLER (VAVD) SERIAL NUMBER(SN): (B)(4). (B)(4).
Additional Manufacturer Narrative
(B)(4). SERVICE REPORT (B)(4) WAS GENERATED FOR THIS EVENT. THE STAMP AND CYLINDER WERE FOUND TO BE DEFECTIVE AND WERE REPLACED. MAINTENANCE WAS PERFORMED PER THE SERVICE PROTOCOL. (B)(4).