inforMED
MalfunctionKDP

VAVD CONTROLLER

Received Jan 29, 2015 · Event occurred Apr 25, 2012

Report 8010762-2014-01035 · MDR key 4461736

Device

Generic name

Regulator, Vacuum

Model number

70103.5905

Catalog number

70103.5905

Product problems

  • Device Operates Differently Than Expected

Patient

Not reported

  • No Information

Narrative

Description of Event or Problem

ON (B)(6) 2012 AT (B)(6) IT WAS REPORTED THAT THE VACUUM CONTROLLER (VAVD) SERIAL NUMBER(SN): (B)(4) DID NOT FUNCTION AND THE POINTER WOULD NOT GO TO ZERO. (B)(4).

Additional Manufacturer Narrative

MAQUET MEDICAL SYSTEM, USA SUBMITS THIS REPORT ON BEHALF OF THE LEGAL MANUFACTURER OF THE DEVICE MAQUET CARDIOPULMONARY (B)(4). SERVICE REPORT (B)(4) WAS GENERATED FOR THIS EVENT. THE MEMBRANE WAS FOUND TO BE DAMAGED AND WAS REPLACED. THE DEVICE WAS TESTED PER THE SERVICE PROTOCOL. (B)(4).