inforMED
MalfunctionKDP

VAVD CONTROLLER

Received Jan 29, 2015 · Event occurred Jan 18, 2012

Report 8010762-2014-01032 · MDR key 4461733

Device

Generic name

Regulator, Vacuum

Model number

70102.4459

Catalog number

70102.4459

Product problems

  • Unstable
  • Device Operates Differently Than Expected

Patient

Not reported

  • No Patient Involvement

Narrative

Description of Event or Problem

ON (B)(6) 2012 THE SSU AT MAQUET DENMARK REPORTED THAT THE VACUUM WAS UNSTABLE, BUT THE MANOMETER WAS FUNCTIONING ON THE VACUUM CONTROLLER (VAVD) SERIAL NUMBER (SN): (B)(4). THE CUSTOMER CHANGED OUT THE MEMBRANE BUT THAT DID NOT RESOLVE THE ISSUE. (B)(4).

Additional Manufacturer Narrative

MAQUET MEDICAL SYSTEM, USA SUBMITS THIS REPORT ON BEHALF OF THE LEGAL MANUFACTURER OF THE DEVICE MAQUET CARDIOPULMONARY (B)(4). SERVICE REPORT (B)(4) WAS GENERATED FOR THIS EVENT. IT WAS DETERMINED THAT THE UNIT WAS REMOUNTED IN THE INCORRECT WAY. THE MOUNTING WAS CORRECTED AND THE DEVICE WAS TESTED PER THE SERVICE PROTOCOL. (B)(4).