FRU, HEADNECKSPINE COIL 1.5T
Received Jan 20, 2015 · Event occurred Dec 30, 2014
Report 1056069-2015-00003 · MDR key 4448177
Device
Product problems
- Device Operates Differently Than Expected
Patient
40 YR
- Erythema
- Burning Sensation
- Burning Sensation
- Partial thickness (Second Degree) Burn
- Partial thickness (Second Degree) Burn
Narrative
Additional Manufacturer Narrative
(B)(4). METHOD: THE INVESTIGATION IS STILL ONGOING ON THIS EVENT. WHEN THE INVESTIGATION IS COMPLETED A FOLLOW UP REPORT WILL BE SENT TO THE FDA. RESULTS: THE INVESTIGATION IS STILL ONGOING ON THIS EVENT. WHEN THE INVESTIGATION IS COMPLETED A FOLLOW UP REPORT WILL BE SENT TO THE FDA. CONCLUSION: THE INVESTIGATION IS STILL ONGOING ON THIS EVENT. WHEN THE INVESTIGATION IS COMPLETED A FOLLOW UP REPORT WILL BE SENT TO THE FDA.
Additional Manufacturer Narrative
(B)(4). BASED ON THE PROVIDED INFORMATION AND TEST PERFORMED ON SITE THERE IS NO INDICATION OF A MALFUNCTION OF THE MR SYSTEM OR COIL. IT IS CONCLUDED THAT THIS HEATING INCIDENT IS A SKIN TO SKIN HEATING INCIDENT CAUSED BY A RF CURRENT LOOP FORMATION INSIDE THE LEFT ARM PIT OF THE PATIENT. NO PADDING WAS USED BETWEEN THE ARMS AND SIDE OF THE PATIENT TO PREVENT SKIN-SKIN CONTACT. IN THIS CASE SEVERAL FACTORS WERE OBSERVED THAT CONTRIBUTED TO THE SUSTAINED INJURY: THE PATIENT WAS OBESE AND COVERED BY A COTTON SHEET, 4 SCANS ON HIGH SAR VALUE WERE CONDUCTED AND THE TOTAL DELIVERED SPECIFIC ENERGY DOSE (SED) WAS 3726 J/KG.
Description of Event or Problem
A MALE PT WAS POSITIONED HEAD FIRST SUPINE AND SCANNED WITH THE DS NVC COIL FOR A CERVICAL SPINE EXAMINATION. THE PT EXPERIENCED A HEATING SENSATION AND AFTER THE EXAMINATION REDDENING OF THE SKIN AND 2 BLISTERS OF 3/4 INCH WERE OBSERVED IN THE PTS LEFT ARM PIT.
Remedial action
- Replace