  • Charred
  • Temperature Problem

Patient

Not reported

  • No Information

Narrative

Additional Manufacturer Narrative

REPORTING DUE TO A FDA FINDING DURING AN AUDIT IN (B)(4) /2014. NO INJURY TO THE PATIENT. CONCLUSION: THE Q2 AND Q26 COMPONENTS WERE THE LIKELY CAUSE OF THE FAILURE, POSSIBLY CAUSED SOME INTERNAL SHORT IN THE BOARD.

Description of Event or Problem

THE UNIT SMELLED LIKE IT WAS BURNING.

Remedial action

  • Replace