SUPPORT ARM 177

Received Dec 4, 2014 · Event occurred Oct 28, 2014

Report 3008355164-2014-00303 · MDR key 4420294

Device

Catalog number

6481720

Product problems

  • Mechanical Problem
  • Unintended Movement

Patient

Not reported

  • Extubate

Narrative

Description of Event or Problem

IT WAS REPORTED THA A PT ACCIDENTALLY WAS EXTUBATED DUE TO A SELF-LOWERING OF THE SUPPORT ARM, THAT HOLDS THE PT'S CIRCUIT, AT THE LEVEL OF JOINTS. THIS INCIDENT LEAD TO AN EMERGENCY RE-INTUBATION, WITH DIFFICULT CONDITIONS. THE RAPID RE-INTUBATION WAS COMPLICATED BY MASSIVE ASPIRATION WITH HEMATOSIS DEGRADATION, REQUIRING OPTIMIZATION OF MECHANICAL VENTILATION, SEDATION, CURARIZATION, AND SWITCHING TO VENTRAL DECUBITUS POSITION (SEVERAL SESSIONS OVER 4/5 DAYS). THE FINAL PT OUTCOME WAS NO INJURY. THE SUPPORT ARM JOINTS WERE TIGHTENED WHICH SOLVED THE PROBLEM AND THE SUPPORT ARM WAS PUT BACK IN USE. (B)(4). MFR REF # 8010042-2014-00549.