inforMED
MalfunctionMOU

GALILEO CENTERING CATHETER III

Received Jan 30, 2003 · Event occurred Dec 5, 2002

Report 2024168-2003-00034 · MDR key 440956

Device

Generic name

Coronary Centering Catheter

Model number

NA

Catalog number

1009491

Lot number

2110851

Product problems

  • Inflation Problem
  • Leak/Splash
  • Other (for use when an appropriate device code cannot be identified)
  • Device, removal of (non-implant)

Patient

UNKNOWN

  • No Consequences Or Impact To Patient
  • Unknown (for use when the patient's condition is not known)

Narrative

Description of Event or Problem

THE BALLOON WAS NOT INFLATING AND THERE WAS A LEAK SOMEWHERE. ATTEMPTS TO INCREASE THE PRESSURE DID NOT RESULT IN A BALLOON INFLATION. AFTER THE DEVICE WAS REMOVED FROM THE PT, FURTHER ATTEMPTS TO INFLATE THE BALLOON WERE MADE WITHOUT SUCCESS. THIS IS BEING FILED BASED ON RETURNED PRODUCT ANALYSIS, WHICH IDENTIFIED A DISTAL SHAFT INFLATION PROXIMAL TO THE BALLOON.