inforMED
MalfunctionJFP

ISE 9180

Received Jan 6, 2015 · Event occurred Dec 12, 2014

Report 1823260-2015-00058 · MDR key 4386781

Device

Generic name

Blood Gas Analyzer

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

05110394001

Lot number

NA

Product problems

  • Low Test Results

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

THERE WERE NO ADVERSE EVENTS. ONLY THE 123 MMOL/L RESULT WAS REPORTED OUTSIDE THE LABORATORY.

Additional Manufacturer Narrative

THE EVENT TOOK PLACE IN (B)(6).

Description of Event or Problem

THE CUSTOMER COMPLAINED OF A QUESTIONABLE ION SELECTIVE ELECTRODE SODIUM RESULT. THE INITIAL RESULT WAS 90 MMOL/L. THE REPEAT RESULT WAS 123 MMOL/L. INFORMATION REGARDING WHICH RESULTS WERE REPORTED OUTSIDE THE LABORATORY AND WHETHER THE PATIENT WAS HARMED WAS REQUESTED BUT NOT PROVIDED. THE LOT NUMBER AND EXPIRATION DATE OF THE SODIUM ELECTRODE WERE REQUESTED BUT NOT PROVIDED. THE FIELD SERVICE REPRESENTATIVE DETERMINED THE ANALYZER WAS NOT APPROPRIATELY GROUNDED. AFTER THE CUSTOMER CORRECTED THE GROUNDING ISSUE, SAMPLES WERE TESTED AND THE RESULTS WERE ACCEPTABLE. THE FIELD SERVICE REPRESENTATIVE DETERMINED THE ANALYZER WAS WORKING PROPERLY.