inforMED
MalfunctionKLC

SPIRIT 1700 DENTAL CHAIR

Received Dec 19, 2014 · Event occurred Dec 3, 2014

Report 1017522-2014-00016 · MDR key 4360995

Device

Generic name

Dental Chair

Manufacturer

Pelton & Crane

Model number

SP17

Catalog number

E1704

Product problems

  • Break

Patient

50 YR

  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

UPON EVALUATION IT WAS IDENTIFIED THAT THE TWO WELDED SUPPORT BLOCK ON THE FRONT TRUSS ASSEMBLY HAD FAILED AS A RESULT OF IMPROPER WELDING FROM OUR SUPPLIER AS IT WAS NOT WELDED TO MANUFACTURERS SPECIFICATIONS. PELTON & CRANE HAS ALREADY CONTACTED THE SUPPLIER OF THE FRONT TRUSS ASSEMBLY AND INFORMED THEN OF THIS ISSUE. THE SUPPLIER INFORMED PELTON & CRANE THAT THEY WILL UPDATE FQC DOCUMENT TO INCLUDE A CONFORMITY CHECK OF WELDMENTS TO ENSURE FRONT TRUSS ASSEMBLIES ARE MANUFACTURED TO PELTON & CRANE SPECIFICATIONS.

Description of Event or Problem

A DENTAL PROFESSIONAL WAS RECLINING A PATIENT FOR TREATMENT IN A PELTON & CRANE SP17 DENTAL CHAIR WHEN THE LIFT ASSEMBLY BROKE RESULTING IN THE SEAT ASSEMBLY TO FALL BACKWARDS TOWARDS THE FLOOR. THE PATIENT DID NOT FALL OUT OF THE CHAIR. THERE WERE NO INJURIES REPORTED.