inforMED
MalfunctionHAW

NDI PASSIVE SPHERES

Received Dec 3, 2014 · Event occurred Oct 16, 2014

Report 3002743211-2014-00045 · MDR key 4346765

Device

Generic name

Reflective Passive Marker Spheres, Haw, Olo

Model number

8801075

Lot number

1407141

Product problems

  • Device Operates Differently Than Expected

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

STERILE SPHERE WERE NOT DETECTED BY CAMERA DURING PROCEDURE. ANOTHER SET OF STERILE SPHERES WERE USED TO COMPLETE THE PROCEDURE WITHOUT ANY FURTHER ISSUES NOTED. THIS EVENT WAS COMMUNICATED BY MEDTRONIC NAVIGATION. SITE/USER DISPOSED OF DEVICE AND DID NOT TAKE ANY PICTURES OF THE DEVICE. CONTRACT MANUFACTURER EVALUATED DHR BASED ON LOT NUMBER. NO QUALITY ISSUES WERE NOTED DURING IN PROCESS INSPECTIONS. NO SERIOUS IMPLICATIONS TO THIS ISSUE WERE MENTIONED BY THE COMPLAINANT. NO PATIENT WAS HARMED AND NO SERIOUS INJURY OCCURRED.