inforMED
MalfunctionGIO

CAPILLARY CAPS

Received Dec 9, 2014 · Event occurred Oct 19, 2014

Report 1217157-2014-00188 · MDR key 4309718

Device

Generic name

Capillary Caps

Catalog number

08685906

Lot number

08685906

Product problems

  • Material Too Rigid or Stiff

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

ADDITIONAL INFORMATION: SIEMENS INVESTIGATION IDENTIFIED THAT A MATERIAL CHANGE MADE BY THE SUPPLIER HAS RESULTED IN CAPILLARY CAPS THAT ARE MORE DIFFICULT TO APPLY TO THE CAPILLARIES. AN URGENT FIELD SAFETY NOTICE #32320 WAS SENT BY E-MAIL TO ALL AFFECTED SIEMENS HEALTHCARE DIAGNOSTICS REGIONAL OFFICES ON JANUARY 22, 2015 BOTH IN THE UNITED STATES AND ELSEWHERE FOR COMMUNICATION WITH AFFECTED CUSTOMERS. PLEASE NOTE: THESE CAPILLARY CAPS ARE AN "OFF THE SHELF" ITEM. SIEMENS PURCHASES THEM FROM A DISTRIBUTOR AND REPACKAGES THEM.

Description of Event or Problem

CUSTOMER REPORTED THAT CAPILLARY CAPS WERE VERY STIFF. CUSTOMER ALSO REPORTED THAT THEY WERE USING THESE CAPS IN COMBINATION WITH SIEMENS 100 & 140 UL GLASS CAPILLARIES. THERE WAS NO REPORT OF INJURY DUE TO THIS EVENT.

Additional Manufacturer Narrative

THIS EVENT IS UNDER INVESTIGATION. THE CAUSE FOR THE EVENT IS UNKNOWN.

Remedial action

  • Recall