inforMED
MalfunctionJAF

THD SLIDE DOPPLER PROBE 5 SHOTS NON STERILE

Received Oct 30, 2014 · Event occurred Sep 24, 2014

Report 3006680097-2014-00004 · MDR key 4261598

Device

Generic name

Doppler Probe Non Sterile

Manufacturer

Thd Spa

Model number

800056B

Catalog number

NA

Lot number

NA

Product problems

  • Communication or Transmission Problem

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

THE DOCTOR OPENED A NEW THD OPERATION BOX, TOOK OUT A NEW THD DOPPLER PROBE AND CONNECTED IT TO THE THD EVOLUTION DEVICE. THE PROBE DID NOT GIVE ANY SIGNAL DURING OPERATION, IT ONLY HISSED. THE DOCTOR RESTARTED THD EVOLUTION, BUT THE PROBLEM (HISS) REMAINED. THE DOCTOR TOOK ANOTHER THD DOPPLER PROBE (FROM ANOTHER BOX) AND COMPLETED OPERATION SUCCESSFULLY.

Additional Manufacturer Narrative

WE ARE WAITING TO RECEIVE THE DEVICE AT THE HEADQUARTERS OF (B)(6), IN ORDER TO COMPLETE THE INVESTIGATION.

Remedial action

  • Notification