THE BAG II DISPOSABLE RESUSCITATOR

Received Oct 31, 2014 · Event occurred Oct 2, 2014

Report 9610483-2014-00003 · MDR key 4236622

Device

Generic name

Manual Disposable Resuscitator

Model number

ADULT

Catalog number

845041

Lot number

040314

Product problems

  • Use of Device Problem
  • Blocked Connection
  • Device Handling Problem

Patient

7 YR

  • Overinflation of Lung
  • Low Oxygen Saturation

Narrative

Description of Event or Problem

LAERDAL MEDICAL (B)(4) HAS INFORMED LAERDAL MEDICAL CORPORATION (LMC) IN NEW YORK OF ANY EVENT WHICH OCCURRED IN (B)(6). ON (B)(6), CHILD WAS ADMITTED TO THE HOSPITAL'S PICU AND AFTER A SUDDEN DROP IN SPO2 WAS INTUBATED AND VENTILATED. A NEW THE BAG II DISPOSABLE RESUSCITATOR WAS CONNECTED TO OXYGEN (>16L/MIN FLOW) AND ATTACHED TO THE PATIENT'S ET TUBE. A NURSE VERIFIED THE RESUSCITATOR SETUP WAS INITIALLY OK. THE DOCTOR HEARD THE EXCESS OXYGEN VENTING OUT THE BACK OF THE RESUSCITATOR AND THE NURSE TOUCHED THE OPEN O2 OVERPRESSURE UMBRELLA VALVE THEREBY BLOCKING THE OXYGEN EXHAUST PORT. DURING THIS USER BLOCKAGE EVENT, THE RESUSCITATOR BAG AND PATIENT AIRWAY BECAME PRESSURIZED BY THE OXYGEN. THE DOCTOR NOTICED THE PATIENT'S LUNGS WERE INFLATED AND REMOVED THE RESUSCITATOR TO RELIEVE THIS PRESSURE. NO REPORTS OF INJURY TO THIS CHILD WERE RECEIVED AND THEIR FINAL OUTCOME IS UNKNOWN.

Additional Manufacturer Narrative

THE CUSTOMER DISPOSED OF THIS RESUSCITATOR AND IT WAS NOT AVAILABLE FOR RETURN TO THE MANUFACTURER. THE HOSPITAL STAFF WAS ONLY ABLE TO RECREATE THIS OVERPRESSURE CONDITION ON OTHER THE BAG IIS USING A SEALED TEST LUNG WITH >15 L/MIN FLOW OF OXYGEN AND THEN MOMENTARILY BLOCKING THE O2 OVERPRESSURE VALVE. AT LOWER O2 FLOWS THE OVERPRESSURE CONDITION WOULD NOT OCCUR. SQUEEZING THE RESUSCITATOR BAG WOULD RELEASE THIS PRESSURE AND RETURN THE BAG II RESUSCITATOR TO NORMAL OPERATION. NO FAULTS WERE FOUND WITH THE BAG II RESUSCITATORS EITHER BEFORE OR AFTER THIS USER BLOCKAGE EVENT. THE O2 OVERPRESSURE VALVE WAS NOT STICKING OR BINDING IN A CLOSED POSITION - THE USER WAS INTENTIONALLY PRESSING THE O2 OVERPRESSURE VALVE CLOSED. THE RESUSCITATOR DFU STATES AN OXYGEN FLOW OF 10 L/MIN IS SUFFICIENT FOR >99% OXYGEN CONCENTRATION IN AN ADULT. BY DESIGN, THE EXTERNAL O2 OVERPRESSURE VALVE IS RECESSED AND PROTECTED FROM CASUAL BLOCKAGES. NO CORRECTIVE ACTION IS RECOMMENDED.