inforMED
InjuryLSL

APTIMA VAGINAL SWAB SPECIMEN COLLECTION KIT

Received Oct 29, 2014 · Event occurred Sep 25, 2014

Report 2024800-2014-00004 · MDR key 4211640

Device

Generic name

In-vitro Diagnostic Assay

Manufacturer

Hologic, Inc.

Model number

301162

Catalog number

301162

Lot number

108908A

Product problems

  • Unintended Ejection
  • Material Fragmentation
  • Improper or Incorrect Procedure or Method

Patient

Not reported

  • Eye Injury
  • Distress

Narrative

Description of Event or Problem

GP HAVING TAKEN NORMAL VAGINAL SWAB WAS TRANSFERRING TO THE TRANSPORT TUBE AND "BREAKING" THE HEAD WITHIN THE TUBE RATHER THAN BREAKING IN ITS DESIGNED SPOT, THE PINK HANDLE BROKE IN TWO PLACES WITH A ONE CM PART FLYING BACK UP WITH SOME FORCE INTO THE GP'S EYE CAUSING DISTRESS TO HER. MAIN PART OF THE SWAB SAMPLE REMAINED IN BOTTLE, MAIN PART OF STEM IN HAND. THE USER TOOK STEPS TO IRRIGATE EYE IN CASE OF LOCAL CONTAMINATION.