Received Oct 9, 2014 · Event occurred Aug 27, 2014

Report 1419185-2014-00019 · MDR key 4181766

Device

Generic name

Vacuum Regulator

Model number

8700-0000-801

Lot number

ISU137460

Product problems

  • Material Integrity Problem

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

COMPLAINT WAS RECEIVED REGARDING WARRANTY REPAIR. UPON RETURN AND SERVICE AND REPAIR INVESTIGATION IT WAS NOTED THAT THE DIAPHRAGM WAS TORN. THE DEVICE IS A VACUUM REGULATOR INTENDED TO REGULATE SUCTION TO REMOVE BODY FLUIDS. IT HAS BEEN DETERMINED THAT A TEAR IN THE REGULATOR DIAPHRAGM WHILE IN USE (AFTER VACUUM HAS BEEN SET) COULD RESULT IN AN INCREASE IN VACUUM TO THE PATIENT THAN WHAT HAD BEEN ORIGINALLY SET.

Additional Manufacturer Narrative

DEVICE IS A VACUUM REGULATOR. THE DEVICE ACHIEVES ITS FUNCTION VIA A REGULATOR MODULE THAT IS EQUIPPED WITH A DIAPHRAGM. THE USER FILED A COMPLAINT THAT THE REGULATOR NEEDED CALIBRATION. DURING THE INVESTIGATION OF THE RETURNED DEVICES IT WAS DISCOVERED THAT THE MALFUNCTIONING UNITS HAD THEIR DIAPHRAGMS TORN. THE USER DID NOT REPORT ANY PATIENT INVOLVEMENT NOR ANY ADVERSE EVENT. THE USER ALSO DID NOT FILE A FORM FDA-3500A. THE MANUFACTURER IS FILING FORM FDA-3500A BECAUSE THIS WAS A MALFUNCTION THAT IF IT WERE TO RECUR, IT WOULD HAVE THE POTENTIAL TO CAUSE SOME PATIENT HARM. INVESTIGATION INTO THE ROOT CAUSE IN ONGOING AS OF THE DATE OF THIS REPORT AND IT HAS NOT YET BEEN DETERMINED THE EXTENT TO WHICH REMEDIAL ACTION WOULD BE APPROPRIATE TO PREVENT AN UNREASONABLE RISK OF SUBSTANTIAL HARM TO PUBLIC HEALTH.

Remedial action

  • Repair