HYBRID CAPTURE 2 RAPID CAPTURE SYSTEM
Received Sep 13, 2002 · Event occurred May 29, 2002
Report 1122376-2002-00001 · MDR key 417453
Device
Generic name
Robotic Plate Processor For Hc2 Ct/gc Dna Test
Manufacturer
Digene Corp.Model number
9000192Catalog number
6000-3101
Lot number
NA
Product problems
- Air Leak
- Bubble(s)
- Inaccurate Delivery
Patient
NA
- No Patient Involvement
Narrative
Description of Event or Problem
THE HYBRID CAPTURE 2 (HC2) RAPID CAPTURE SYSTEM (RCS) IS A SEMI-AUTOMATED INSTRUMENT APPLICATION OF THE HC2 CT/GC, CT-ID AND GC-ID DNA TESTS. THE APPLICATION USES A ROBOTIC MICROPLATE PROCESSOR THAT AUTOMATES THE PIPETTING AND MICROPLATE HANDLING STEPS OF THE HC2 TESTS. COMPLAINT REPORTED THAT 3 SPECIMENS OF 40 WERE NOT TRANSFERRED IN TRAINING CONDUCTED AT A NEW CUSTOMER SITE. TWO MISSED SPECIMENS WERE TRAINING PANEL SIMULATED SPECIMENS. ONE WAS A DIGENE CERVICAL SAMPLER WITHOUT PATIENT SPECIMEN. THE REPORT INDICATED THAT THE RCS FUNCTIONED AS EXPECTED. THE MISSED TRANSFERS WERE IDENTIFIED PRIOR TO PROCESSING IN THE ROUTINE TRANSFER VERIFICATION STEP, AS DESCRIBED IN THE APPLICATION MANUAL INSTRUCTION, AND WERE CORRECTED BY MANUAL PIPETTING BEFORE PROCESSING. THE REPORT NOTED AIR POCKETS, ALSO REFERRED TO AS 'AIR BUBBLES", IN THE SPECIMENS AFTER VORTEXING. INITIAL REVIEW CONCLUDED THAT THE REPORT DID NOT PERTAIN TO OR HAVE THE POTENTIAL TO CAUSE INJURY. NO PATIENT SPECIMENS WERE INVOLVED. SIMULATED SPECIMENS WERE IN USE FOR TRAINING PURPOSES.
Remedial action
- Modification/Adjustment