inforMED
MalfunctionHAW

NDI PASSIVE SPHERES

Received Jul 11, 2014 · Event occurred Jan 21, 2011

Report 3002743211-2014-00001 · MDR key 4170682

Device

Generic name

Reflective Passive Marker Sphere - Haw, Olo

Lot number

4797801

Product problems

  • Device Misassembled During Manufacturing /Shipping

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

OPERATING ROOM STAFF AT (B)(6) NOTED A PASSIVE MARKER SPHERE HAD DELAMINATED REFLECTIVE FILM. THIS WAS NOTED DURING A PROCEDURE. AS SPHERE WAS USED DURING A PROCEDURE IT WAS DESTROYED VIA AUTOCLAVE AS IT WAS USED IN A PROCEDURE AND CONSIDERED A BIOHAZARD. DURING INVESTIGATION IT WAS DETERMINED THAT THE ADHESIVE FOR THE FILM WAS IMPROPERLY APPLIED. THIS ISSUE WAS CORRECTED AND THIS FAILURE HAS NOT RECURRED SINCE THIS REPORT. NO SERIOUS IMPLICATIONS TO THIS ISSUE WERE MENTIONED BY THE COMPLAINANT. NO PT WAS HARMED AND NO SERIOUS INJURY OCCURRED.