NDI PASSIVE SPHERES

Received Jul 11, 2014 · Event occurred May 9, 2014

Report 3002743211-2014-00010 · MDR key 4170612

Device

Generic name

Reflective Passive Marker Spheres

Model number

8801075

Lot number

1308271

Product problems

  • Device Operates Differently Than Expected

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

REFLECTIVE STERILE SPHERES FAILED TO TRACK DURING A PROCEDURE. ANOTHER SET OF STERILE SPHERES WERE USED TO COMPLETE THE PROCEDURE. THIS ISSUE WAS COMMUNICATED BY MEDTRONIC NAVIGATION. SPHERES WERE ANALYZED BY MEDTRONIC NAVIGATION AND REPORTED AS HAVING A "BLOTCHY, IRREGULAR SURFACE". FURTHER ANALYSIS AT NORTHERN DIGITAL INC. WAS NOT COMPLETED AS SPHERES WERE USED IN A PROCEDURE AND CONSIDERED A BIOHAZARD. DECONTAMINATION OF SPHERE WOULD RENDER INVESTIGATION IMPOSSIBLE. SITE/USER DID NOT TAKE PICTURES OF SPHERE. REVIEW OF DHRS SHOWED NO QUALITY ISSUES. CONTRACT MFR (IIMAK) UPDATED PROCEDURES TO INSPECT FOR "BLOTCHY, IRREGULAR SPHERES". HOWEVER, SINCE THIS IS THE ONLY KNOWN OCCURRENCE OF THIS FAILURE, NO ROOT CAUSE CAN BE ESTABLISHED. NO SERIOUS IMPLICATIONS TO THIS ISSUE WERE MENTIONED BY THE COMPLAINANT. NO PT WAS HARMED AND NO SERIOUS INJURY OCCURRED.