SYNTHES DRILL BIT (ON LOAN)

Received Aug 27, 1992 · Event occurred Feb 18, 1992

Report 4158 · MDR key 4158

Device

Patient

54 YR

Narrative

Description of Event or Problem

POST OPERATIVE X-RAY DEMONSTRATED FOREIGN BODY. (PIECE OF DRILLBIT DEEP IN THE BONE). INSTRUMENT INSPECTION REVEALED TIP OF DRILL BIT MISSING. FURTHER TREATMENT WAS NOT INDICATED ACCORDING TO ATTENDING PODIATRIST.DEVICE LABELED FOR SINGLE USE. PATIENT MEDICAL STATUS PRIOR TO EVENT: SATISFACTORY CONDITION. THERE WAS NOT MULTIPLE PATIENT INVOLVEMENT.INVALID DATA - ON DEVICE SERVICE/MAINTENANCE. NO DATA - REGARDING DATE LAST SERVICED. SERVICE PROVIDED BY: INVALID DATA. INVALID DATA - SERVICE RECORDS AVAILABILITY. NO IMMINENT HAZARD TO PUBLIC HEALTH CLAIMED. DEVICE USED AS LABELED/INTENDED.DEVICE WAS EVALUATED AFTER THE EVENT. METHOD OF EVALUATION: ACTUAL DEVICE INVOLVED IN INCIDENT WAS EVALUATED, VISUAL EXAMINATION. RESULTS OF EVALUATION: NONE OR UNKNOWN. CONCLUSION: DEVICE FAILURE DIRECTLY CAUSED EVENT. CERTAINTY OF DEVICE AS CAUSE OF OR CONTRIBUTOR TO EVENT: INVALID DATA. CORRECTIVE ACTIONS: DEVICE PERMANENTLY REMOVED FROM SERVICE, DEVICE RETURNED TO MANUFACTURER/DEALER/DISTRIBUTOR. THE DEVICE WAS NOT DESTROYED/DISPOSED OF.