inforMED
InjuryFTH

VERION DIGITAL MARKER M FOR ZEISS MICROSCOPES

Received Aug 1, 2014 · Event occurred Jun 4, 2014

Report 3010300699-2014-00002 · MDR key 4024527

Device

Generic name

Camera, Still, Microsurgical

Model number

X-SPM

Catalog number

8065998244

Lot number

NA

Product problems

  • Display or Visual Feedback Problem

Patient

73 YR

  • Postoperative refraction, unexpected

Narrative

Description of Event or Problem

AN OPHTHALMIC SURGEON REPORTED THAT AFTER CATARACT SURGERY SHE IDENTIFIED A DEVIATION IN THE AXIS FOR THE IMPLANTED TORIC IOL IN THREE PATIENTS. DURING SURGERY, SHE SAW A ROTATION OF THE FINAL AXIS COMPARED WITH THE INITIALLY PLANNED AXIS. THIS LED TO RESIDUAL ASTIGMATISM THAT WAS CORRECTED WITH EYE GLASSES. FOR THIS PT SURGEON REPORTED A JUMP ON THE MARKS FOR THE INCISION OF 30 DEGREES. THE SURGEON INDICATED THE POSTOPERATIVE ASTIGMATISM WAS DUE TO THE SYSTEM MAKINGS AND NOT DUE TO A POSTOPERATIVE IOL ROTATION. ADDITIONAL INFO WAS RECEIVED FROM THE SURGEON. DURING SURGERY, MARKINGS IN THE MICROSCOPE AND IN THE DISPLAY WERE PERMANENTLY CHANGING, THUS MAKING IT DIFFICULT TO GUESS WHICH WAS THE CORRECT TORIC ALIGNMENT AXIS. THE SURGEON CHOSE ONE OF THE MULTIPLE AXIS DISPLAYED AND POSTOPERATIVELY THE IOL PRESENTED A ROTATION OF 12 DEGREES DEVIATION FROM THE PREOPERATORY AXIS SUGGESTED BY THE DIGITAL MARKER (164 DEGREE INSTEAD OF 176 DEGREE). THE ROTATION WAS OBSERVED ON THE FIRST POSTOPERATIVE DAY AND PERSISTED ONE MONTH AFTER SURGERY. THE PT WAS NOT WILLING TO UNDERGO IOL REPOSITIONING. ADDITIONAL INFO HAS BEEN REQUESTED BUT NOT RECEIVED TO DATE.

Additional Manufacturer Narrative

INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFO BECOMES AVAILABLE. ADDITIONAL INFO HAS BEEN REQUESTED BUT NOT RECEIVED TO DATE. (B)(4).