MalfunctionMRP

INOMAX DSIR (DELIVERY SYSTEM)

Received Jul 16, 2014

Report 3004531588-2014-00032 · MDR key 4013960

Device

Generic name

Apparatus, Nitric Oxide Delivery

Manufacturer

Ikaria

Model number

10007

Product problems

  • Failure to Sense

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

FAILED NO SENSOR (DEVICE ISSUE). NO ADVERSE EVENT. CASE DESCRIPTION: THIS INITIAL NON-SERIOUS DEVICE CASE REPORT WAS RECEIVED ON (B)(4) 2014 FROM A RESPIRATORY THERAPIST (RT) IN THE UNITED STATES WHO CALLED IKARIA TECHNICAL SERVICES (TS) REGARDING A DEVICE ISSUE WITH INOMAX DSIR (B)(4). THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE DEVICE ISSUE AND THE REPORTER STATED THAT NO HARM OCCURRED TO THE PATIENT. IT WAS REPORTED THAT THE RT CALLED TO REPORT A DEVICE ISSUE WITH INOMAX DSIR (B)(4) WHILE ON A PATIENT AND HAD EXPERIENCED AN ALARM CONDITION FAILED NO SENSOR. THE SET UP WAS REVIEWED WITH THE REPORTER AND FOUND TO BE CORRECT. ALL CONNECTIONS WERE ALSO CHECKED AND FOUND TO BE TIGHT. A HIGH RANGE CALIBRATION TO NITRIC OXIDE FUEL CELL FOLLOWED BY A LOW RANGE CALIBRATION WAS PERFORMED ON THE DEVICE WHICH RESULTED IN A PREVIOUS ALARM CONDITION MESSAGE FAILED NO SENSOR. THE REPORTER STATED NO HARM OCCURRED TO THE PATIENT. INOMAX DSIR (B)(4) WAS REMOVED FROM SERVICE AND RETURNED TO IKARIA FOR SERVICE EVAL. INVESTIGATIONAL RESULTS WERE RECEIVED ON (B)(4) 2014. ON (B)(4) 2014: THIS DEVICE CASE DID NOT RESULT IN AN ADVERSE EVENT; HOWEVER, IT IS BEING REPORTED BECAUSE A SIMILAR DEVICE ISSUE OCCURRED IN THE PAST THAT RESULTED IN A SERIOUS ADVERSE EVENT (REFER TO MDR # 3004531588-2013-00022).

Additional Manufacturer Narrative

ON (B)(6) 2014 A RESPIRATORY THERAPIST (RT) CALLED IKARIA TECHNICAL SERVICES (TS) REGARDING A DEVICE ISSUE WITH INOMAX DSIR (B)(4). THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE DEVICE ISSUE AND NO HARM OCCURRED TO THE PATIENT ((B)(4)). DEVICE INVESTIGATION WAS COMPLETED ON (B)(4) 2014. INOMAX DSIR (B)(4) WAS RETURNED TO THE MFR FOR SERVICE INVESTIGATION. THE IKARIA REGIONAL SERVICE CENTER (RSC) REVIEWED THE SERVICE LOG AND FINDINGS CONFIRMED THE REPORTED COMPLAINT OF A FAILED NO SENSOR ALARM. THE FAILED NO SENSOR ALARM IMMEDIATELY FOLLOWED A FAILED LOW NO CALIBRATION WITH LOW POINT: ((B)(4)) COUNTS ABOVE THE ALLOWED MAXIMUM OF (B)(4). A FURTHER LOG REVIEW REVEALED A DELIVERY FAILURE (DF) ALARM FOR NO > 100 PPM WHILE DELIVERING 5 PPM NO WITH AVG=4.5 CUR=1.5 LPM IM FLOW AND DELIVERING 0 PPM NO WITH AVG=5.4 CUR=1.4 LPM IM FLOW. BOTH DF ALARMS OCCURRED LESS THAN 9 MINUTES BEFORE THE FAILED NO SENSOR ALARM AND ARE CONSISTENT WITH A MISBEHAVING NO CELL. THE RSC INVESTIGATION ALSO EXPERIENCED A FAILED NO SENSOR ALARM AT BOOTUP, PERFORMED LOW AND HIGH CALIBRATIONS TO CLEAR THE ALARM AND REPLACED THE NO SENSOR AS A PRECAUTIONARY MEASURE. A FULL FUNCTIONAL TEST WAS PERFORMED AND THE DEVICE OPERATED ACCORDING TO SPECIFICATIONS SO IT WAS RETURNED TO THE DEVICE SERVICE POOL. THE ROOT CAUSE FOR THIS INCIDENT WAS NO CALIBRATION LOW COUNTS ABOVE MAXIMUM. THIS CONDITION WILL BE TRACKED AND TRENDED UNDER IKARIA'S QUALITY SYSTEM. THIS CASE DID NOT RESULT IN AN ADVERSE EVENT; HOWEVER, IT IS BEING SUBMITTED TO REG AUTHORITIES BECAUSE A SIMILAR FAILURE OCCURRED IN THE PAST WHICH RESULTED IN A SERIOUS ADVERSE EVENT (MDR 3004531588-2013-00022).