Received Jul 3, 2014 · Event occurred May 23, 2014

Report 2020676-2014-00025 · MDR key 3983290

Device

Generic name

Hyperbaric Monoplace Chamber

Model number

3300H

Catalog number

3300H-01-1

Product problems

  • Device Operates Differently Than Expected

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

DURING CLINICAL USE THE CHAMBER DECOMPRESSED FROM 2.0 AT TO 0 ATA IN 6.5 MINUTES. TESTING OF THE CHAMBER WAS DONE, BY THE CENTER STAFF, AFTER THE PT WAS REMOVED AND THE FAILURE WAS DUPLICATED. INFO PROVIDED BY THE CENTER STATES THAT THE PT WAS NOT INJURED. THEY ARE REPORTED AS DOING "FINE".

Additional Manufacturer Narrative

THE REPORTED COMPLAINT THAT THE CHAMBER IS DECOMPRESSING FASTER THAN SPEC WAS VERIFIED. THE DECOMPRESSION TIMING FROM 2.0 ATA TO 0 ATA IN 6.5 MINUTES COULD BE A POTENTIAL SAFETY ISSUE IF THE PT HAS LUNG DISEASE AND HOLDS THEIR BREATH. THIS COULD RESULT IN GAS BECOMING TRAPPED IN THEIR CHEST. THE CHAMBER WAS EVALUATED BY A TECHNICIAN FROM THE MFR. HE FOUND THAT THE RATE VALVE WS OUT OF CALIBRATION AND REQUIRED AN ADJUSTMENT. A NOTE WAS INCLUDED IN THE REPORT STATING THAT THE ISSUE WAS RESOLVED. THE CAUSE OF THE RATE VALVE GOING OUT OF SPEC AND INCREASING THE DECOMPRESSION TIMING IS UNK. THE PROBABLE CAUSE IS DRIFT.