inforMED
InjuryEMA

MAXCEM ELITE

Received Jul 30, 2014

Report 2024312-2014-00545 · MDR key 3967961

Device

Generic name

Cement, Dental

Manufacturer

Kerr Corporation

Catalog number

33872

Product problems

  • Loss of or Failure to Bond

Patient

21 YR

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

A DOCTOR ALLEGED THAT A PATIENT HAD EXPERIENCED THE DEBONDING OF A VENEER APPROXIMATELY ONE (1) MONTH AFTER PLACEMENT WITH THE MAXCEM ELITE CLEAR PRODUCT.

Additional Manufacturer Narrative

MAXCEM ELITE INSTRUCTIONS FOR USE STATE: "FOR VENEER CEMENTATION USING MAXCEM ELITE, ADHESIVE APPLICATION ON TOOTH PREP IS REQUIRED." THE DOCTOR REPORTED THAT AN ADHESIVE HAD NOT BEEN APPLIED TO THE TOOTH PRIOR TO INITIAL CEMENTATION. THE PATIENT HAD RETURNED TO THE OFFICE ON (B)(4) 2014 AND THE VENEER WAS CLEANED OUT AND RE-CEMENTED USING MAXCEM ELITE ALONG WITH AN ADHESIVE APPLICATION. TO DATE, THE PATIENT IS DOING FINE. THE PRODUCT INVOLVED IN THE ALLEGED INCIDENT WAS NOT RETURNED AND NO LOT NUMBER WAS PROVIDED; THEREFORE, NO EVALUATION CAN BE CONDUCTED.