inforMED
MalfunctionFPL

VOYAGER

Received Jul 17, 2014 · Event occurred Apr 8, 2014

Report 3965835 · MDR key 3965835

Device

Generic name

Incubator, Neonatal Transport

Model number

VOYAGER

Catalog number

N/A

Lot number

N/A

Product problems

  • Failure to Charge
  • Loose or Intermittent Connection
  • Smoking

Patient

*

  • No Information

Narrative

Description of Event or Problem

WHILE REMOVING OLD BATTERY RELATED TO FAILURE TO CHARGE BECAUSE OF OPEN MAIN FUSE, AS TERMINALS WERE LOOSENED, WIRING HARNESS SHIFTED AND WENT UP IN SMOKE.BIOMED REVIEWED. SHORT FROM POSITIVE BATTERY TERMINAL TO GROUND CAUSED QUICK DISCHARGE OF CURRENT. CONTACT WAS POSSIBLY MADE BY TECH'S WRENCH. TECH SAYS YES IT DID DUE TO HIS REACTION OF THE WIRE GOING UP IN SMOKE.