A CORD
Received Jun 13, 2014 · Event occurred May 16, 2014
Report 8010047-2014-00310 · MDR key 3958438
Device
Generic name
A Cord
Manufacturer
Olympus Medical Systems CorporationModel number
MH-969Catalog number
NA
Lot number
UNK
Product problems
- Break
Patient
Not reported
- No Consequences Or Impact To Patient
Narrative
Description of Event or Problem
OLYMPUS MEDICAL SYSTEMS CORP (OMSC) WAS INFORMED THAT DURING POLYPECTOMY OF COLON THE DOCTOR GOT AN ERROR TRYING TO ACTIVATE OUTPUT WHILE HE SNARED THE TARGET POLYP AND OUTPUT COULD NOT BE ACTIVATED . THEN HE NOTICED THAT THE SUBJECT PRODUCT BROKE. HE EXCHANGED THE SUBJECT PRODUCT AND THE ELECTROSURGICAL UNIT FOR OTHER DEVICES AND COMPLETED THE PROCEDURE. THERE WAS NO REPAIR OF PATIENT INJURY REGARDING THIS REPORT.
Additional Manufacturer Narrative
THE SUBJECT DEVICE WAS RETURNED TO OMSC FOR INVESTIGATION. THE INVESTIGATION CONFIRMED THAT THE CABLE BROKE NEAR A CODE PLUG. THE CABLE CORE WAS BURNT AND THE BROKEN OUT SECTION WAS DISCOLORED. OMSC WAS INFORMED THAT THE USER FACILITY HAD USED THE SUBJECT DEVICE FOR NEARLY 10 YEARS. OMSC ASSUMES THAT BENDING LOAD WAS PUT ON THE CABLE REPEATEDLY OVER LONG PERIODS AND IT CAUSED THE BREAKAGE OF THE CABLE CORE. IN ADDITION, THE USER FACILITY CONTINUED TO USE THE SUBJECT DEVICE WITH THE CABLE CORE BROKEN, IT CAUSED THIS EVENT. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVIC REPORT IN AN ABUNDANCE OF CAUTION.