LAERDAL SILICON RESUSCITATOR (LSR)

Received Jun 25, 2014 · Event occurred May 1, 2014

Report 9610483-2014-00002 · MDR key 3918863

Device

Generic name

Manual Reusable Resuscitator

Model number

ADULT

Catalog number

870000

Lot number

080403

Product problems

  • Use of Device Problem
  • Adverse Event Without Identified Device or Use Problem
  • Device Handling Problem

Patient

51 YR

  • Cardiac Arrest
  • Pneumothorax
  • Low Oxygen Saturation

Narrative

Description of Event or Problem

LAERDAL MEDICAL A.S. (LMAS) IN (B)(6) HAS INFORMED LAERDAL MEDICAL CORPORATION (LMC) IN (B)(6) OF AN EVENT WHICH OCCURRED IN (B)(6). A FEMALE HOSPITAL PATIENT IN PREPARATION FOR A THYROIDECTOMY OPERATION BEGAN COUGHING. SHE WAS BEING MANUALLY VENTILATED WITH OXYGEN THROUGH A LAERDAL SILICONE RESUSCITATOR (LSR) WHILE THE ANESTHESIA MACHINE WAS PREPARED. HER OXYGEN SATURATION DECREASED OVER A 2 MINUTE TIME PERIOD AND SHE WENT INTO ASYSTOLE. THE LSR WAS REMOVED FROM THE PATIENT AND A LARGE AIR FLOW / EXHALE OCCURRED. CPR WAS PROVIDED FOR ABOUT ONE MINUTE AND CIRCULATION RETURNED. A THORAX AND ABDOMEN CT SCAN CONFIRMED PNEUMOTHORAX IN BOTH LUNGS. THE PATIENT SURVIVED THIS EVENT.

Additional Manufacturer Narrative

THIS LSR WAS RETURNED TO LMAS AND IT PASSED ALL INSPECTIONS AND FUNCTIONAL TESTS AND MEETS ALL SPECIFICATIONS. AN OVER PRESSURE CONDITION COULD BE CREATED IF OXYGEN WAS CONNECTED TO THE LSR AND IT HAD THE OLD STYLE OF LSR INTAKE VALVE (PRE 2003); NO EXTERNAL LSR RESERVOIR VALVE WAS CONNECTED; AND THE 23 MM OD LSR INTAKE PORT THEN BLOCKED/OCCLUDED. THIS OVER PRESSURE CONDITION COULD BE RELIEVED BY DELIVERING ONE, OR MORE, LSR VENTILATIONS OR OPENING/VENTING THE BAG/AIRWAY CONDUIT. THE LSR DFU RECOMMENDS USING THE LSR RESERVOIR VALVE AND OXYGEN RESERVOIR TO DELIVER HIGHER OXYGEN CONCENTRATIONS. THE LSR OXYGEN RESERVOIR WILL NOT CONNECT TO THE LSR WITHOUT THE RESERVOIR VALVE. WITH THE LSR RESERVOIR VALVE CONNECTED THIS CONDITION COULD NOT HAVE OCCURRED. THE CURRENT LSR CONFIGURATION (SINCE 2003) IS NOT SUSCEPTIBLE TO THIS UNUSUAL USER ERROR. UNUSUAL OCCURRENCE. WE HAVE LITTLE DETAIL OF USER'S EQUIPMENT CONFIGURATIONS DURING THIS EVENT.