inforMED
MalfunctionNVV

SUNQUEST POWERPATH

Received Mar 6, 2014 · Event occurred Feb 6, 2014

Report 2029302-2014-00002 · MDR key 3906855

Device

Generic name

Powerpath

Product problems

  • Data Problem

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

NINETY-ONE SITES HAVE BEEN NOTIFIED VIA PRODUCT SAFETY NOTICE (PSN) AND A CORRECTION WILL BE AVAILABLE FOR 10.0 AND 10.0.1 AND WILL BE CONTAINED IN ALL FUTURE VERSIONS. THERE ARE 0 SITES HAVE ALREADY BEEN CORRECTED. STEPS TO REPRODUCE: CREATE A NEW CASE. MANUALLY ENTER THE HIS ORDER NUMBER FOR AN IMPORTED ORDER. SAVE THE CASE.

Description of Event or Problem

WHEN AN ORDER IS RECEIVED IN POWERPATH FROM A SENDING FACILITY, IF THE SENDING FACILITY SENDS THE SAME ORDER NUMBER FOR DIFFERENT PATIENTS, POWERPATH DOES NOT VERIFY THE ORDER NUMBER WHICH COULD RESULT IN A DUPLICATE ORDER NUMBER FROM THE SAME FACILITY.

Remedial action

  • Modification/Adjustment