SUNQUEST POWERPATH
Received Mar 6, 2014 · Event occurred Feb 6, 2014
Report 2029302-2014-00002 · MDR key 3906855
Device
Generic name
Powerpath
Manufacturer
Sunquest Info SystemsProduct problems
- Data Problem
Patient
Not reported
- No Consequences Or Impact To Patient
Narrative
Additional Manufacturer Narrative
NINETY-ONE SITES HAVE BEEN NOTIFIED VIA PRODUCT SAFETY NOTICE (PSN) AND A CORRECTION WILL BE AVAILABLE FOR 10.0 AND 10.0.1 AND WILL BE CONTAINED IN ALL FUTURE VERSIONS. THERE ARE 0 SITES HAVE ALREADY BEEN CORRECTED. STEPS TO REPRODUCE: CREATE A NEW CASE. MANUALLY ENTER THE HIS ORDER NUMBER FOR AN IMPORTED ORDER. SAVE THE CASE.
Description of Event or Problem
WHEN AN ORDER IS RECEIVED IN POWERPATH FROM A SENDING FACILITY, IF THE SENDING FACILITY SENDS THE SAME ORDER NUMBER FOR DIFFERENT PATIENTS, POWERPATH DOES NOT VERIFY THE ORDER NUMBER WHICH COULD RESULT IN A DUPLICATE ORDER NUMBER FROM THE SAME FACILITY.
Remedial action
- Modification/Adjustment