inforMED
MalfunctionIPF

PENA MUSCLE STIMULATOR PROBE (NOT STERILE)

Received Feb 24, 2014

Report 9612007-2014-00010 · MDR key 3902047

Device

Generic name

Pena Muscle Stimulator Probe

Catalog number

PS-2P

Product problems

  • Device Operates Differently Than Expected
  • Scratched Material

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

THE THIRD REPORT OF EIGHT FROM THE SAME FACILITY INVOLVING PENA PROBE. CROSS REFERENCE WITH MFR. REPORT 9612007-2014-00003. A REPORT WAS RECEIVED FROM A CHILDREN'S HOSPITAL THAT THE PENA PROBE STOPPED/DIDN'T WORK. THE CONSULTANT SURGEON CUT THE CABLE AND EXPOSED THE INNER CABLING WHICH APPEARED TO BE OXIDIZED. THE PROBE WAS IN CONTACT WITH A PATIENT (AGE OF THE PATIENT IS NOT AVAILABLE). THERE WAS A DELAY WHILE THE STAFF LOCATED ANOTHER PROBE TO USE. THERE WAS NO INJURY INVOLVED AS A RESULT OF THE DELAY.

Additional Manufacturer Narrative

THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS BEEN RECEIVED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED UPON THE REPORTED INFORMATION.