ACORN 120 STRAIGH STAIRLIFT
Received Apr 2, 2014 · Event occurred Sep 25, 2013
Report 3003124453-2014-00002 · MDR key 3901686
Device
Generic name
Powered Stairway Chair Lift
Manufacturer
Acorn Mobility Services LtdModel number
ACORN 120Catalog number
NA
Lot number
NA
Product problems
- Insufficient Device Problem Information
Patient
79 YR
- Death
Narrative
Description of Event or Problem
ACORN MOBILITY SERVICES LTD ((B)(4)) THE MFG CORPORATION OF THE MEDICAL DEVICE CONCERNED, WAS, IN (B)(4) 2013, CONTACTED BY (B)(6) TO PROVIDE ASSISTANCE FOR THEIR INVESTIGATIONS INTO A FATAL ACCIDENT IN (B)(6) INVOLVING THE ACORN 120 STRAIGHT STAIRLIFT. (B)(4) WERE INFORMED THAT THE INDIVIDUAL USING THE DEVICE HAD A HISTORY OF HEALTH PROBLEMS. FOLLOWING THIS CORRESPONDENCE, TWO MECHANICAL EXPERTS CONDUCTED AN INSPECTION OF THE STAIRLIFT ON (B)(4) 2013. THE EXPERTS OBSERVED THAT THE HORIZONTAL WELD SECURING THE SEATPOST BASE TO THE MAIN SEAT SUPPORT ASSEMBLY OF THE STAIRLIFT USED BY THIS INDIVIDUAL HAD FAILED. AFTER THIS INCIDENT, ACORN (B)(4) CONDUCTED INDEPENDENT TESTS OF SEATPOSTS FROM SIMILAR STAIRLIFT AND THE TEST RESULTS HAVE REVEALED NO ISSUES REGARDING POTENTIAL FAILURES OF SEATPOSTS. NEVERTHELESS, AS A RESULT OF THIS EVENT AND THE EXISTENCE OF ONE ADD'L INCIDENT ((B)(6)), ACORN (B)(4) HAS, AS A PRECAUTIONARY MEASURE, DECIDED TO IMPLEMENT A SEATPOST REPLACEMENT PROGRAM ON A (B)(4) BASIS. THIS IS DESPITE THERE BEING NO OFFICIAL OR CONFIRMED LINK ON THE CAUSES OF FAILURE FOR THE TWO INCIDENTS, THE LIMITED INFO ACORN (B)(4) HAS REGARDING THE INCIDENTS AND THE POSITIVE TEST RESULTS RELATED TO THE STRUCTURAL INTEGRITY OF SIMILAR STAIRLIFT PRODUCTS. ACORN (B)(4) IS NOT AWARE OF ANY INCIDENTS INVOLVING FAILURE OF THE HORIZONTAL WELD SECURING THE SEATPOST BASE TO THE MAIN SEAT SUPPORT ASSEMBLY OF THE STAIRLIFT ANYWHERE OTHER THAN (B)(6). THE TYPE OF SEATPOSTS POTENTIALLY IMPLICATED IN THESE INCIDENTS WERE (B)(4).