inforMED
MalfunctionFSO

OPMI PICO ON S100 FLOOR STAND

Received Feb 6, 2014 · Event occurred Jan 8, 2014

Report 9615010-2014-00006 · MDR key 3894155

Device

Generic name

Microscope, Surgical, General And Pla

Model number

S100 FLOORSTAND

Catalog number

000000-1403-542

Lot number

NA

Product problems

  • Installation-Related Problem

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE CUSTOMER HEARD A NOISE AND WENT INTO THE PROCEDURE ROOM. THEY FOUND THE LAMP HOUSING (XENON POWER SUPPLY ILLUMINATION UNIT) HAD FALLEN AND WAS HANGING BY THE CABLES FROM THEIR OPMI PICO MICROSCOPE S100 FLOOR STAND. THE CUSTOMER CONFIRMED THAT NO ONE WAS INJURED.

Additional Manufacturer Narrative

THE FIELD SERVICE ENGINEER (FSE) FOUND THAT ALL FOUR BOLTS SECURING THE XENON POWER SUPPLY ILLUMINATION UNIT MOUNTING PLATE TO THE S100 STAND WERE BROKEN. THE MICROSCOPE WAS REPAIRED. SITE CONTACT IS SAME AS INITIAL REPORTER.

Remedial action

  • Repair