Device

Generic name

Radiologic Table

Manufacturer

Siemens Ag

Model number

00475509

Product problems

  • Unintended System Motion

Patient

97 YR

  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

THE OPERATOR MANUAL FOR MULTIX TOP SYSTEM (AXB1-150.620.01.01.02) CONTAINS WARNINGS REGARDING COLLISION/CRUSHING ZONES. THE REPORTED ISSUE IS UNDER INVESTIGATION AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION HAS BEEN RECEIVED. THIS REPORT WAS SUBMITTED (B)(4) 2013.

Description of Event or Problem

IT WAS REPORTED THAT A (B)(6) YEAR OLD MALE PT WAS PRESENTED TO THE FACILITY FOR AN X-RAY OF A HAND. THE PT WAS SITTING IN A WHEELCHAIR ON THE SIDE OF THE MULTIX TOP SYSTEM TABLE WITH HIS LEGS UNDER THE TABLE. THE OPERATOR WAS BEHIND THE FOLDING SCREEN WHEN THE TABLE STARTED DESCENDING ON THE PT'S THIGHS WITHOUT ANY GIVEN COMMANDS. THE OPERATOR STOPPED THE DOWNWARD MOVEMENT BY PRESSING THE RED EMERGENCY STOP BUTTON. THE PT SUSTAINED NO INJURY AND DID NOT REQUIRE ANY MEDICAL ATTENTION. THE REPORTED EVENT OCCURRED IN (B)(6).