inforMED
MalfunctionDHX

ADVIA CENTAUR XP CEA ASSAY

Received May 16, 2014 · Event occurred Apr 24, 2014

Report 1219913-2014-00124 · MDR key 3815868

Device

Generic name

Cea Immunoassay,, Product Code:

Model number

N/A

Catalog number

09788458

Lot number

154

Product problems

  • False Reading From Device Non-Compliance
  • High Test Results

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

A FALSELY ELEVATED ADVIA CENTAUR XP CEA RESULT WAS OBTAINED BY THE CUSTOMER ON A PATIENT SAMPLE. THE PATIENT'S CEA RESULT WAS HISTORICALLY LOWER AND THE CUSTOMER PERFORMED REPEAT TESTING. THE CEA RESULTS OF THE REPEAT TESTING WERE LOWER. THE PATIENT SAMPLE WAS ALSO RUN ON ANOTHER ADVIA CENTAUR SYSTEM FOR VERIFICATION AND THE RESULT WAS LOW. THE LOW ADVIA CENTAUR XP RESULT WAS REPORTED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED ADVIA CENTAUR XP CEA RESULT.

Additional Manufacturer Narrative

THE CAUSE FOR THE INITIAL FALSELY ELEVATED ADVIA CENTAUR XP CEA RESULT WHEN COMPARED TO THE LOWER REPEAT RESULTS AND THE PATIENT'S HISTORICAL CEA TEST RESULT IS UNKNOWN. THE CUSTOMER'S QUALITY CONTROL RESULTS WERE WITHIN ACCEPTABLE RANGE LIMITS. NO CONCLUSION CAN BE DRAWN. THE INSTRUCTION FOR USE (IFU) LIMITATION SECTION STATES: "NOTE: DO NOT INTERPRET LEVELS OF CEA AS ABSOLUTE EVIDENCE OF THE PRESENCE OR THE ABSENCE OF MALIGNANT DISEASE. MEASUREMENTS OF CEA SHOULD ALWAYS BE USED IN CONJUNCTION WITH OTHER DIAGNOSTIC PROCEDURES, INCLUDING INFORMATION FROM THE PATIENT'S CLINICAL EVALUATION. THE CONCENTRATION OF CEA IN A GIVEN SPECIMEN DETERMINED WITH ASSAYS FROM DIFFERENT MANUFACTURERS CAN VARY DUE TO DIFFERENCES IN ASSAY METHODS, CALIBRATION, AND REAGENT SPECIFICITY. CEA DETERMINED WITH DIFFERENT MANUFACTURERS' ASSAYS WILL VARY DEPENDING ON THE METHOD OF STANDARDIZATION AND ANTIBODY SPECIFICITY. WARNING: DO NOT USE THE ADVIA CENTAUR CEA IMMUNOASSAY AS A SCREENING TEST FOR DIAGNOSIS." THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS.

Remedial action

  • Inspection