inforMED
MalfunctionEBC

TOOTH CONDITIONER GEL

Received Jan 22, 2002 · Event occurred Jan 16, 2002

Report 2515379-2002-00004 · MDR key 373968

Device

Generic name

Dental Restorative Material

Manufacturer

L.d. Caulk

Model number

NA

Catalog number

646125

Lot number

010531

Product problems

  • Detachment Of Device Component

Patient

UNKNOWN

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

THE COMPLAINT COULD NOT BE REPRODUCED WHEN RETURN SAMPLES (WHERE PROVIDED) AND/OR RETAIN SAMPLES WERE EVALUATED. THEREFORE, CO DID NOT POSITIVELY IDENTIFY ANY FAILURE MODE OR MECHANISM.

Description of Event or Problem

TIPS WERE ATTACHED TO THE SYRINGES AND THE SYRINGES WERE THEN EXTRUDED. THERE IS NO FORMALLY DOCUMENTED METHOD FOR THIS ECTRUSION TEST; IT IS GENERALLY ASSOCIATED WITH APPEARANCE METHOD GM-062-89.

Description of Event or Problem

SINCE THE ALLEGED DEFECT COULD NOT BE REPRODUCED INTERNALLY, CO'S CONCLUSION IS THAT THE DOCTOR DID NOT PROPERLY ATTACH THE TIP TO THE FILLED SYRINGE. ALSO, THE DIRECTIONS FOR USE STATE "GEL SHOULD FLOW FREELY WITH GENTLE PRESSURE. DO NOT USE EXCESSIVE FORCE. IF NOT , REMOVE SYRINGE FROM PT FIELD AND CHECK FOR OBSTRUCTION." FAILURE TO FOLLOW THESE INSTRUCTIONS COULD LEAD TO THE COMPLAINTS OF TIPS COMING OFF. THE DIRECTIONS FOR USE ALSO STATE TO "DISCARD NEEDLES AFTER USE, AS NEEDLES MAY CLOG IF GEL IS ALLOWED TO DRY INSIDE." CLOGGED TIPS COULD CREATE THE NEED FOR EXCESSIVE FORCE, LEADING TO POTENTIAL FAILURE.

Description of Event or Problem

THE DISPENSING TIP CAME OFF OF THE SYRINGE IN THE PATIENT'S MOUTH. THERE WAS NO INJURY.