SHUNT SENSOR SYS500
Received Nov 1, 2013 · Event occurred Oct 23, 2013
Report 1124841-2013-00195 · MDR key 3726667
Device
Generic name
Blood-gas Monitor
Manufacturer
Terumo Cardiovascular Systems Corp.Model number
CDI510HCatalog number
NA
Lot number
QE29L
Product problems
- Fluid/Blood Leak
Patient
UNK
- Blood Loss
Narrative
Description of Event or Problem
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR SYSTEMS THAT DURING CARDIOPULMONARY BYPASS SURGERY THE SHUNT SENSOR LEAKED. LESS THAN 1CC BLOOD LOSS. PRODUCT WAS CHANGED OUT. SURGERY WAS SUCCESSFUL.
Additional Manufacturer Narrative
UPON EVALUATION OF THE DEVICE THE COMPLAINT WAS CONFIRMED. THE UNIT WAS VISUALLY INSPECTED AND PRESSURIZED, AND A LEAK WAS OBSERVED AT THE THERMOWELL. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO ANOMALIES. THIS UNIT WAS SUFFICIENTLY CURED AND SEALED TO PASS THROUGH A 100 PERCENT LEAK TEST. THE ROOT CAUSE, A LEAK FROM THE THERMOWELL POSITION RESULTING FROM STRESS, WAS IDENTIFIED TO BE SHIPPING AND HANDLING PAIRED WITH THE COUPLING OF THE SENSOR TO THE BPM HEAD. (B)(4). ALL AVAILABLE INFORMATION HAS BEEN PLACED ON FILE IN QUALITY MANAGEMENT FOR APPROPRIATE TRACKING, TRENDING AND FOLLOW UP.