OtherFBJ
KARL STORZ
Received Jan 10, 2002 · Event occurred Apr 20, 2001
Report 2020550-2001-00032 · MDR key 371161
Device
Generic name
High Frequency Cord
Manufacturer
Karl Storz Gmbh & Co. KgModel number
277KBCatalog number
277KB
Lot number
CK
Product problems
- Detachment Of Device Component
- Intraprocedure, fire or flash during
- Increase in Pressure
- Sparking
Patient
UNKNOWN
- Injury
- Surgical procedure aborted/stopped
Narrative
Description of Event or Problem
THE CO WAS CONTACTED BY PT'S ATTORNEY WHO REPORTED THAT "A FIRE BROKE OUT ON THE OPERATING TABLE DURING THE USE OF A HYSTEROSCOPE AND RESECTOSCOPE." AFTER CONTACTING HOSP, THE CO WAS TOLD THAT A HIGH FREQUENCY CORD WAS BLOWN FROM THE CONNECTOR AND A SPARK MADE CONTACT WITH INCONTINENCE PAD. PAD CAUGHT ON FIRE CAUSING INJURIES TO THE PT. PROCEDURE CANCELLED.