inforMED
MalfunctionEXR

UNOMETER BAG 3L BO (50/100)

Received Dec 19, 2013 · Event occurred Nov 20, 2013

Report 3007966929-2013-00021 · MDR key 3669798

Device

Generic name

Device, Urine Flow Rate Measuring

Manufacturer

Unomedical Ltd.

Model number

MM84062182

Catalog number

MM84062182

Lot number

652617

Product problems

  • False Device Output
  • Device Markings/Labelling Problem

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

INFORMATION RECEIVED VIA COMPLAINT FORM INDICATES THE FOLLOWING: "SCALE ON THE REPLACEMENT URINE COLLECTING BAG FOR UNOMETER 500 IS NOT REPRESENTING THE ACTUAL CONTENTS. A NURSE NOTICED SIGNIFICANT DEVIATION AND PUT THE BAG TO THE TEST. WITH A SYRINGE, SHE FILLED 250 ML OF WATER INTO IT AND THE SCALE SHOWED APPROXIMATELY 500 ML.

Additional Manufacturer Narrative

BASED ON THE AVAILABLE INFORMATION, THIS EVENT IS DEEMED A REPORTABLE MALFUNCTION. THERE WERE NO REPORTS OF A PT BEING HARMED AS A RESULT OF THIS MALFUNCTION. A REQUEST WAS SENT ON (B)(4) 2013 TO SUPPLIER TO INVESTIGATE THIS ISSUE. NO ADDITIONAL PT/EVENT DETAILS HAVE BEEN PROVIDED TO DATE. A RETURN SAMPLE FOR EVALUATION IS NOT EXPECTED. SHOULD ADDITIONAL INFORMATION BECOMES AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. REPORTED TO THE FDA ON (B)(4) 2013. NOTE: THE ACTUAL DATE OF EVENT IS UNKNOWN, SO THE DATE USED WAS THE DATE CONVATEC BECAME AWARE.