inforMED
MalfunctionDPZ

OXIMAX N-560 PULSE OXIMETER

Received Dec 9, 2013 · Event occurred Nov 11, 2013

Report 2936999-2013-00905 · MDR key 3663100

Device

Generic name

Pulse Oximeter

Manufacturer

Mediana

Model number

N-560

Catalog number

N560

Product problems

  • No Display/Image

Patient

Not reported

  • No Patient Involvement

Narrative

Description of Event or Problem

IT WAS REPORTED TO COVIDIEN THAT THE UNIT COULD NOT DISPLAY THE SPO2 AND HEART RATE VALUE CORRECTLY AND RESULTED IN THE ACTUAL VALUE TO NOT BE RECOGNIZABLE. THERE WAS NO PT INVOLVED.

Additional Manufacturer Narrative

COVIDIEN SERVICE VERIFIED THE MISSING SEGMENTS REPORT. THE FRONT PCB WAS REPLACED. THE MANUFACTURE DATE IS UNK FROM THE SERIAL NUMBER PROVIDED. (B)(4).