inforMED
MalfunctionJJW

ICHEM VELOCITY CONTROL

Received Dec 11, 2013 · Event occurred Aug 31, 2013

Report 2023446-2013-00034 · MDR key 3636136

Device

Generic name

Ichem Velocity Control

Manufacturer

Iris International

Lot number

213-13

Product problems

  • Incorrect Or Inadequate Test Results

Patient

Not reported

  • No Information

Narrative

Description of Event or Problem

CUSTOMER GETTING DISCORDANT RESULTS WITH 20 PATIENTS BETWEEN ICHEM VELOCITY (NEGATIVE) AND IQ200 (POSITIVE). CUSTOMER REPORTED THAT ONE PATIENT SAMPLE WAS NEGATIVE ON CHEMISTRY AND HAD >45 RBC'S. THE OTHER 19 PATIENT SAMPLES WERE ALSO NEGATIVE ON CHEMISTRY AND HAD APPROXIMATELY 5 RBC'S. THE PRESENCE OF RBC'S WAS CONFIRMED WITH MANUAL MICROSCOPE IN ALL PATIENTS.

Additional Manufacturer Narrative

2023446-2013-00034 . IRIS PRODUCT IFU INSERT STATES THAT THE DIAGNOSTIC OR THERAPEUTIC DECISIONS SHOULD NOT BE BASED ON ANY SINGLE RESULT OR METHOD. NO REPORTS OF CHANGE IN PATIENT MANAGEMENT OR INJURIES WERE REPORTED. THE FOLLOWING MDRS ARE RELATED TO THIS EVENT. 2023446-2013-0024, -00025, -00026, -00027, -00028, -00029, -00030, -00031, -00032, -00033, -0034, -00035, -00036, -00037, -00038, -00039, -00040, -00041, -00042.