3M STERI-VAC STERILIZER/AERATOR 5XL
Received Jan 20, 2014 · Event occurred Dec 24, 2013
Report 2110898-2014-00004 · MDR key 3608905
Device
Generic name
Ethylene Oxide Gas Sterilizer
Manufacturer
3m Health CareModel number
5XLCatalog number
5XLP
Lot number
NA
Product problems
- Appropriate Device Problem Term/Code Not Available
Patient
UNK
- Headache
Narrative
Additional Manufacturer Narrative
THERE WERE TWO UNITS THAT WERE PLACED ON EACH OTHER. ONE UPPER AND ONE LOWER. THE UPPER UNIT WAS REPORTED TO BE ON ACCELERATED MODE AND HAD A SERIAL NUMBER OF (B)(4).
Description of Event or Problem
IT WAS REPORTED THAT A MALE EMPLOYEE WITHIN THE HOSPITAL STERILIZATION FACILITY WAS EXPOSED TO EO WHEN A FEMALE OPERATOR OF THE 3M EO STERILIZER OPENED THE DOOR PRIOR TO FULL AERATION AFTER BEING INFORMED TO ABORT CYCLE. IT WAS REPORTED THE STERILIZER WAS ON ACCELERATED MODE WHEN THE EVENT OCCURRED. THE ACCELERATED MODE SHOULD NOT HAVE BEEN USED FOR A STANDARD STERILIZATION PROCESS FOR THIS MODE IS ONLY APPLICABLE WHEN UNIT IS BEING SERVICED. USER ERROR CONTRIBUTED TO THIS EVENT. THE MALE EMPLOYEE SOUGHT MEDICAL CONSULTATION AT THE OCCUPATIONAL HEALTH CLINIC WITHIN THE HOSPITAL. IT WAS REPORTED THAT THE EMPLOYEE EXPERIENCED A HEADACHE. AT THIS TIME, IT IS NOT KNOWN WHETHER ANY MEDICAL INTERVENTION WAS ADMINISTERED FOR THE HOSPITAL WOULD NOT RELEASE THIS INFO. THE ESTIMATED EXPOSURE TIME TO THE OPERATOR WAS 15 TO 30 SECONDS. THERE WAS ANOTHER INDIVIDUAL (FEMALE OPERATOR OF STERILIZER) AND THIS SUBJECT WAS REPORTED UNDER 2110898-2014-00003. THIS REPORT INCLUDES MORE INFO CONCERNING THE ACTUAL EVENT AND SERVICE INFO. THIS REPORT ONLY COVERS THE MALE SUBJECT REGARDING HIS EXPOSURE.