Narrative

Description of Event or Problem

IT WAS REPORTED TO COVIDIEN THAT THE UNIT DISPLAY BEGAN MISSING SEGMENTS IN THE HEART RATE PORTION OF THE DISPLAY. THERE WAS NO PATIENT HARM.

Additional Manufacturer Narrative

COVIDIEN INVESTIGATION VERIFIED THE UNIT STARTED TO LOSE SEGMENTS INTERMITTENTLY. A VISUAL INSPECTION FOUND A COLD SOLDER JOINT ON ONE OF THE 8-SEGMENT DISPLAY POSTS WHICH COULD CAUSE AN INTERMITTENT FAILURE. THE DISPLAY PCB IS REPLACED. (B)(4).