inforMED
MalfunctionIZT

MINIVIDAS

Received Nov 12, 2001 · Event occurred Oct 8, 2001

Report 1950204-2001-00016 · MDR key 360507

Device

Generic name

Immunoflurometric Equipment

Manufacturer

Biomerieux, Inc.

Model number

MINIVIDAS

Catalog number

W2011

Lot number

NA

Product problems

  • Device Issue

Patient

*

  • Unknown (for use when the patient's condition is not known)

Narrative

Description of Event or Problem

CUSTOMER WAS RUNNING "QCT" TEST STRIPS ON PUMP AND FAILED THE TEST 4 TIMES. QCR TEST IS A TEST TO DETECT CLOGGED PUMP TUBES. CLOGGED PUMP TUBES COULD RESULT IN FALSE RESULTS BEING REPORTED TO THE PT. THERE ARE NO REPORTS OF FALSE RESULTS BEING GENERATED AT THIS SITE.

Remedial action

  • Replace
  • Notification
  • Inspection
  • Repair