inforMED
OtherJIO

MULTISTIX 10SG STRIPS

Received Jan 13, 2014 · Event occurred Nov 25, 2013

Report 1217157-2013-00260 · MDR key 3601616

Device

Generic name

Multistix 10sg

Catalog number

10627148

Product problems

  • False Positive Result

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

CUSTOMER REPORTED 27 FALSE POSITIVE UROBILINOGEN RESULTS USING DETERIORATED MULTISTIX 10SG TEST STRIPS. CUSTOMER INDICATED THAT THEY HAVE BEEN RUNNING PATIENT SAMPLES ON THESE TEST STRIPS SINCE (B)(6) 2013. THERE WAS NO REPORT OF INJURY DUE TO THIS EVENT.

Additional Manufacturer Narrative

CUSTOMER INDICATED THAT THEY RUN PATIENT SAMPLES EVEN WHEN THE QUALITY CONTROL (QC) FAILED. CUSTOMER SHOULD NOT HAVE RUN THE PATIENT SAMPLES WHEN QC WAS FAILED. CUSTOMER HAS BEEN PROVIDED WITH NEW BOTTLE OF MULTISTIX 10SG STRIPS. CUSTOMER INDICATED THAT THE ISSUE IS RESOLVED WITH A NEW LOT OF MULTISTIX 10SG STRIPS. THE EVENT HAS OCCURRED DUE TO AN OPERATOR ERROR.