inforMED
MalfunctionOVZ

ORTHOSORB 1 PIN 50X1.3MM KIT

Received Jan 16, 2014 · Event occurred Dec 6, 2013

Report 0001825034-2014-00434 · MDR key 3579727

Device

Generic name

Pin, Fixation

Manufacturer

Biomet Orthopedics

Model number

N/A

Catalog number

841070

Lot number

326136

Product problems

  • Fracture

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. DEVICE AVAILABILITY - THE DEVICE IS REPORTED TO BE AVAILABLE FOR EVALUATION; HOWEVER, IT HAS NOT BEEN RECEIVED BY BIOMET ORTHOPEDICS TO DATE. IN THE EVENT THAT THE DEVICE IS RECEIVED AND EVALUATED, A FOLLOW UP REPORT WILL BE SENT TO THE FDA TO PROVIDE RESULTS. (NOTE: BIOMET, INC. ACQUIRED THE TRAUMA PRODUCT LINE FROM DEPUY ORTHOPAEDICS, INC. (¿DEPUY¿) ON JUNE 16, 2012 (¿CLOSING DATE¿). PURSUANT TO THE WRITTEN AGREEMENT BETWEEN BIOMET AND DEPUY, BIOMET AGREED TO BE RESPONSIBLE FOR REGULATORY REPORTING FOR EVENTS WHICH OCCURRED AFTER THE CLOSING DATE REGARDLESS OF THE ENTITY THAT ACTUALLY MANUFACTURED THE PRODUCT OR ACTUALLY SOLD THE PRODUCT TO THE HEALTHCARE PROVIDER. BECAUSE THE PRODUCT THAT IS THE SUBJECT MATTER WAS MANUFACTURED BEFORE THE CLOSING DATE, PLEASE BE ADVISED THAT THE SUBJECT PRODUCT WAS MANUFACTURED BY DEPUY AND NOT BIOMET.) DEPUY ALSO SOLD THE PRODUCT THAT IS THE SUBJECT MATTER TO THE HEALTHCARE PROVIDER INVOLVED.

Additional Manufacturer Narrative

EXAMINATION OF RETURNED DEVICE FOUND NO EVIDENCE OF PRODUCT NON-CONFORMANCE. EVALUATION OF DEVICE FOUND EVIDENCE THAT FAILURE MODE WAS LIKELY DUE TO MISUSE BY IMPLANT BEING PUT THROUGH EXCESSIVE FORCE OR OVER TORQUED, AND/OR NOT INSPECTED FOR WEAR AND DISFIGUREMENT WHICH MAY HAVE PREVENTED THE USE OF THE IMPLANT AND ITS FAILURE.

Description of Event or Problem

IT WAS REPORTED THAT PATIENT UNDERWENT A BUNNIONECTOMY UTILIZING A PIN ON (B)(6) 2013. DURING THE PROCEDURE, THE PIN FRACTURED UPON INSERTION. ANOTHER PIN WAS UTILIZED TO COMPLETE THE PROCEDURE WITH NO INJURY TO THE PATIENT.