inforMED
MalfunctionJPA

STA COMPACT

Received Dec 26, 2013 · Event occurred Oct 14, 2013

Report 2245451-2013-00002 · MDR key 3560380

Device

Catalog number

58986

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

THE CUSTOMER DISCOVERED THAT THE LOT SPECIFIC INTERNATIONAL SENSITIVITY INDEX (ISI) VALUE DISPLAYED ON ONE OF HIS SIA COMPACT ANALYZERS WAS RESET TO 1.00 INSTEAD OF 1.32. THIS DEFECT WAS DUE TO AN OPERATOR ERROR AND RESULTED TO THE UNDERESTIMATION OF THE INTERNATIONAL NORMALIZED RATIO (INR) VALUE FOR 203 PATIENTS WHO HAD A PROLONGED PT TEST RESULT (INR > 1.5). THE CLINICAL PHARMACIST IDENTIFIED SOME PATIENTS WHEREIN DOSAGE OF TREATMENT HAD BEEN ADJUSTED BASED ON THOSE ERRONEOUS RESULTS. HOWEVER, ACCORDING TO THE CUSTOMER, NO DEATH OR SERIOUS INJURY HAS BEEN REPORTED TO DATE. REFERENCE MFR REPORT NUMBER: 8043723-2013-00002.