FORCEPS, INSERT, PLATE SHAPED
Received Dec 27, 2013 · Event occurred Dec 1, 2013
Report 1418479-2013-00036 · MDR key 3554805
Device
Generic name
Forceps Insert
Manufacturer
Richard Wolf GmbhModel number
8393.182Catalog number
8393.182
Lot number
1130672
Product problems
- Break
Patient
Not reported
- No Information
Narrative
Additional Manufacturer Narrative
AN INVESTIGATION WAS COMPLETED AS THE ACTUAL DEVICE WAS RETURNED TO THE RWMIC FACILITY ON (B)(6) 2013. SEGMENT CONNECTING PULLROD WITH JAW WAS BROKEN. IN ADDITION, BOTTOM HALF OF WALL SURROUNDING THE PIN WAS MISSING. TEETH SHOW SIGNS OF HEAVY USE. WHEN FORCEPS ASSEMBLED WITH SHEATH, ALL PARTS ARE CONTAINED. DEVICE HISTORY: MANUFACTURED 05/2011, SOLD ON (B)(4) 2012; NO RECORD OF ANY REPAIRS OR ROUTINE MAINTENANCE OF DEVICE. NO SIMILAR COMPLAINTS IN THE LAST FIVE YEARS. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. I.E. INTENDED USE, INDICATIONS AND FIELD OF USE, PREPARATION AND CAUTIONS. RICHARD WOLF CONSIDERS THIS MATTER CLOSED. HOWEVER, IN THE EVENT WE RECEIVE ADDITIONAL INFO, WE WILL PROVIDE FDA WITH FOLLOW-UP INFO.
Description of Event or Problem
SALES REPRESENTATIVE REPORTED TO RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) A PATIENT RETURNED TO HOSPITAL AFTER A BARIATRIC PROCEDURE AND IS CURRENTLY IN FOR A PROLONGED STAY. THERE WERE MANY INSTRUMENTS USED DURING THE PROCEDURE. TWO HAVE BEN IDENTIFIED AS BEING MANUFACTURED BY RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (B)(4). FORCEPS INSERT (8393.182) (B)(4). INSULATED SHEATH (8393.913) (B)(4).