inforMED
InjuryHLN

HUMPHREY INSTRUMENT

Received Aug 24, 2001 · Event occurred Jul 16, 2001

Report 349265 · MDR key 349265

Device

Generic name

Ultrasonic Biometer

Model number

820

Catalog number

820-4152

Lot number

*

Product problems

  • Malfunction

Patient

68 YR

  • Surgical procedure

Narrative

Description of Event or Problem

ON AUGUST 7, 2001, IT WAS DISCOVERED THAT PT WOULD NEED SURGERY TO REPLACE LENS THAT WAS INSERTED IN THE PT'S LEFT EYE. VISION IN LEFT EYE WAS 20/70 PRE OPERATIVE. VISION IN LEFT EYE WAS 20/400. PRE-OPERATIVE AXIAL LENGTH WAS 24.31. RECHECK OF AXIAL LENGTH WAS 22.99. USING THE SAME ULTRASONIC BIOMETER. CALIBRATION WAS CHECKED AND WAS CORRECT.